Monopolar - Computer Controlled intelligent Cut with Feedback Responce System
1 Three Cut modes-(Auto Cut-optional)
(a) Pure cut suitable for under water cutting
(TRUP, TUVP & TCRE) with URO-CUT, which eliminates lag time in intiating cut especially suted for urological procedure.
(b) Three Blend - modes (program)
(c) Endocut (Gastrocut) - for G.I Surgery
2 Three Coag mode- (Auto Coag-Optional)
(a) Desiccation/Soft- for low voltage soft coagulation suitable for laproscopic & delicate tissue work.
(b) Fulgurate / Forced - non contact Coagulation for minimal smoke generation
(c) Spray for non-contact large tissue area coagulation with minimum depth of necrosis. bipolar-SLS (Computer controlled self Limiting System-Unmodulated-Auto Optinal)
Micro- for delicate tissue work by controlled desiccation without charring, adhesion or blancing of adjacent tissue (opthalmic & nuero-surergy).
Macro- for new generation of macro bipolar procedure etc. When used with BIPOL probes provides excellent tamponade capabilities.
Precise- for contorlled and effective vessel sealing. |
|
It deactivates the unit & alarms in case of any discountinuity in the patient plate cord. Exculsive PEM (Patient Electrode monitoring)
Isolated RF output directs current to return back to generator to reduce the risk of Alternate site burns & to ensure patient safety in case of any discontinuity in patient plate cord. This current has no reference to ground so it does not cause notice able interference with the ECG monotor, Cardiac pace maker etc.
Independent activation of hand switching, foot switching & bipolar output to protect against indvertent output from an idle accessory lying on or near the patient or surgical staff.
Class '2' leakage cancelling, a must for lengthy procedure like Cardiac surgery (CABG) etc.
Autoranging PPCQM - patient plate Contact Quality Monitoring technology - monitors the quality and quantity of contact between the patient and the patient palte. If this contact is not proper the unit deactivates & alarms, thus eliminating any risk of Passive electrode site burns. |